<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-11-21T05:03:34Z</responseDate><request verb="GetRecord" identifier="oai:macsphere.mcmaster.ca:11375/16526" metadataPrefix="dim">https://prod-ms-be.lib.mcmaster.ca/server/oai/request</request><GetRecord><record><header><identifier>oai:macsphere.mcmaster.ca:11375/16526</identifier><datestamp>2026-09-24T22:38:36Z</datestamp><setSpec>com_11375_271</setSpec><setSpec>col_11375_272</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor">Ostapiak, Orest</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author">Tunkr, Ayman</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department" lang="en_US">Medical Physics</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2014-12-02T20:11:31Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="available">2014-12-02T20:11:31Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued">2015-06</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">http://hdl.handle.net/11375/16526</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://doi.org/10.71548/14660</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">The aims of this study are first, to assess and compare the amount of&#xd;
deformation for seven thermoplastic masks from three different manufacturers&#xd;
[Aquaplast RT-1889YS (Aqua variable perf), Fiberplast RT-1889KYS (Fiber variable&#xd;
perf), Embrace RT-8320 (Embrace), AccuPerf IMRT head mask RG461-4SHT&#xd;
(AccuPerf), Fiberplast RT-1889KS (Fiberplast), Aquaplast RT-1989S (Aqua thick), and&#xd;
Aquaplast RT-1889S( Aquaplast)] using time lapse photography, and second, to&#xd;
measure the amount of pressure exerted by each mask on a rigid head phantom&#xd;
using pressure sensitive film.&#xd;
For the first aim, all masks were marked with fiducial points located near&#xd;
forehead, nose, and chin. The masks were fabricated on a rigid head phantom&#xd;
following manufacturer instructions. Each mask was mounted on a plastic mounting&#xd;
tray, and photographed in a sagittal view over the span of one week. The fiducial&#xd;
coordinates were determined by a center of mass algorithm. A scale factor was&#xd;
determined by photographing a graph sheet mounted in place of the mask.&#xd;
For the second aim, the pressure exerted by each mask was evaluated by&#xd;
mounting the mask on the head phantom after sandwiching a strip of pressure&#xd;
sensitive film between the two at three anatomical locations: the forehead, nose,&#xd;
and chin.&#xd;
Measurement of translation in the posterior direction did not give&#xd;
reproducible results as direct caliper measurements did not agree with those from&#xd;
the time-lapse photography technique.&#xd;
In the superior/inferior direction, the Embrace mask showed the greatest&#xd;
deformation at forehead and nose areas (1.069±0.017mm (inferiorly), and&#xd;
0.432±0.010mm (inferiorly) respectively), while the Aqua thick mask showed the&#xd;
greatest deformation at chin area (0.970±0.016mm (superiorly)). The averaged median pressure of the masks at the forehead location were&#xd;
not significantly different except for the Fiber variable perf which showed the&#xd;
greatest averaged median pressure ( 0.118 MPa). At the nose location, no mask&#xd;
showed a significantly different averaged median pressure. At the chin location, the&#xd;
Aqua thick mask showed a significantly higher averaged median pressure compared&#xd;
to the others (0.180 MPa).</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degreetype" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree" lang="en_US">Master of Applied Science (MASc)</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso" lang="en_US">en</dim:field>
   <dim:field mdschema="dc" element="title" lang="en_US">Assessment of Various Thermoplastic Masks Used for Radiation Therapy Immobilization in Terms of Shrinkage and Pressure</dim:field>
   <dim:field mdschema="dc" element="type" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="others" element="access-status">open.access</dim:field>
</dim:dim></metadata></record></GetRecord></OAI-PMH>