<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-11-21T01:25:57Z</responseDate><request verb="GetRecord" identifier="oai:macsphere.mcmaster.ca:11375/19908" metadataPrefix="dim">https://prod-ms-be.lib.mcmaster.ca/server/oai/request</request><GetRecord><record><header><identifier>oai:macsphere.mcmaster.ca:11375/19908</identifier><datestamp>2026-09-24T21:34:56Z</datestamp><setSpec>com_11375_271</setSpec><setSpec>col_11375_272</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor">Stern, F.</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author">Piggott, Benjamin Arthur</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department" lang="en_US">Engineering Physics</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2016-07-21T23:32:07Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="available">2016-07-21T23:32:07Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued">1978</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">http://hdl.handle.net/11375/19908</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://doi.org/10.71548/21358</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">&lt;p> The Three Beam Gamma Densitometers developed by AECL-WNRE for measurement of void fraction in flowing steam/water mixtures were investigated to determine the major causes of drift and slow response to changes in void fraction. Available remedies were evaluated. The study indicated the following: (1) Using the existing current mode system, considerable reductions in drift and response time may be possible through the following&#xd;
inexpensive modifications: (a) design changes in the photo multiplier tube high voltage&#xd;
circuit,&#xd;
(b) replacement of the detectors with Cadmium Telluride semi-conductor detectors,&#xd;
(c) replacement of the scintillation phosphor with one which exhibits no afterglow such as Bismuth Germanate or Cesium Fluoride,&#xd;
(d) matching voltages of photo multiplier tube outside wall and photocathode,&#xd;
(e) improved stabilization of high voltage and detector temperature. (2) Drift would probably be reduced to less than 0.1%/hour if the existing current mode system were replaced with a simple integral counting mode system at an approximate cost of $5,200 per densitometer. (3) Additions to the above counting mode system of a single channel analyser (S.C.A.) and an automatic gain control unit which uses a reference light would probably completely eliminate drift as well as substantially improve system response. This would require an additional $4,900 per densitometer. &lt;/p> &lt;p> It is recommended that these modifications be implemented in the same order as above, on a trial basis, as far as further expenditure is justified.&lt;/p></dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degreetype" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree" lang="en_US">Master of Engineering (MEngr)</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso" lang="en_US">en_US</dim:field>
   <dim:field mdschema="dc" element="subject" lang="en_US">gamma densitometers, Atomic Energy, Canada Limited, performance, temperature, drift</dim:field>
   <dim:field mdschema="dc" element="title" lang="en_US">Evaluation of the Performance of the Gamma Densitometers of Atomic Energy of Canada Limited (Part A)</dim:field>
   <dim:field mdschema="dc" element="type" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="others" element="access-status">open.access</dim:field>