<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-11-21T04:41:14Z</responseDate><request verb="GetRecord" identifier="oai:macsphere.mcmaster.ca:11375/21428" metadataPrefix="dim">https://prod-ms-be.lib.mcmaster.ca/server/oai/request</request><GetRecord><record><header><identifier>oai:macsphere.mcmaster.ca:11375/21428</identifier><datestamp>2026-09-24T02:28:27Z</datestamp><setSpec>com_11375_271</setSpec><setSpec>col_11375_272</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor">Byun, Soo-Hyun</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author">Anjomani, Zahra</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department" lang="en_US">Medical Physics</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2017-05-17T17:14:22Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="available">2017-05-17T17:14:22Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued">2017</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">http://hdl.handle.net/11375/21428</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://doi.org/10.71548/22911</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">Tissue Equivalent Proportional Counters (TEPCs) are considered the standard instrument&#xd;
for microdosimetry, aiming at measuring the distribution of the energy deposited&#xd;
by ionizing radiation in a micrometric target, and have been employed for a number&#xd;
of radiation physics, radiation protection and radiation biology applications.&#xd;
This study describes development procedure and performance of a novel multi-element&#xd;
TEPC. Following an extensive Geant4 simulation study, an advanced prototype&#xd;
multi-element gaseous microdosimetric detector was developed using the Thick Gas&#xd;
Electron Multiplier (THGEM) technique. The multi-element design was employed&#xd;
to increase the neutron detection efficiency. The prototype THGEM multi-element&#xd;
detector consists of three alternating layers of tissue equivalent plastic hexagons and&#xd;
each layer houses a hexagonal array of seven cylindrical gas cavity elements with equal&#xd;
heights and diameters of 17 mm. The final detector structure incorporates 21 gaseous&#xd;
volumes. Owing to the absence of wire electrodes, the THGEM multi-element detector&#xd;
offers flexible and convenient fabrication in contrast to the traditional wire-based&#xd;
methods. The detector responses to neutron and gamma-ray were investigated using&#xd;
the McMaster Tandetron 7Li(p,n) neutron source. The dosimetric performance of the&#xd;
detector is presented in contrast to the response of a commercial tissue equivalent&#xd;
proportional counter. The collected spectra exhibit the expected features of the lineal&#xd;
energy distributions for given proton beam energies. Compared to the 2 inch TEPC&#xd;
response, the detector gave a consistent microdosimetric response with a maximum&#xd;
discrepancy of 15% in measured neutron absorbed dose. An improvement of a factor&#xd;
of 3.0 in neutron detection efficiency has been accomplished. The prototype detector&#xd;
offers a simple fabrication process and provides the fundamental basis for development&#xd;
of a high efficiency TEPC dedicated for monitoring weak neutron radiation fields.</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degreetype" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree" lang="en_US">Doctor of Philosophy (PhD)</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso" lang="en_US">en</dim:field>
   <dim:field mdschema="dc" element="title" lang="en_US">DEVELOPMENT OF A THICK GAS ELECTRON MULTIPLIER-BASED MULTI-ELEMENT MICRODOSIMETRIC DETECTOR</dim:field>
   <dim:field mdschema="dc" element="type" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="others" element="access-status">open.access</dim:field>
</dim:dim></metadata></record></GetRecord></OAI-PMH>