<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-11-21T03:45:35Z</responseDate><request verb="GetRecord" identifier="oai:macsphere.mcmaster.ca:11375/26261" metadataPrefix="dim">https://prod-ms-be.lib.mcmaster.ca/server/oai/request</request><GetRecord><record><header><identifier>oai:macsphere.mcmaster.ca:11375/26261</identifier><datestamp>2026-09-24T20:48:54Z</datestamp><setSpec>com_11375_271</setSpec><setSpec>col_11375_272</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor">Farncombe, Troy</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author">Reimers, Erik</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department" lang="en_US">Biomedical Engineering</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2021-03-18T20:16:37Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="available">2021-03-18T20:16:37Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued">2021</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">http://hdl.handle.net/11375/26261</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://doi.org/10.71548/25344</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract" lang="en_US">The current forefront of research in medical imaging technology includes working towards the combination of multiple imaging modalities into a single scan. SPECT is an&#xd;
imaging modality that would benefit from a combined approach with MRI allowing for&#xd;
improved diagnostic capabilities particularly in soft tissue regions. However, the traditional detectors used in SPECT are not capable of operating in the presence of an MRI&#xd;
magnetic field. This has led to the requirement for development of new gamma camera&#xd;
designs using modern solid state detectors.&#xd;
We have developed the hardware and software for a gamma camera design using&#xd;
cadmium zinc telluride detector technology. The camera will be used in a SPECT&#xd;
system being made up of multiple individual gamma camera modules. These modules&#xd;
will be placed in a ring that can be inserted into the MRI bore as a stationary insert&#xd;
for simultaneous SPECT/MR imaging. Testing of the mutual compatibility of the two&#xd;
systems has led to the development of a gradient triggering hardware and software&#xd;
protocol. Subsequent testing of this method in a clinical 3T MR was performed. When&#xd;
gradient triggering was enabled, erroneous event data from the mutual interference was&#xd;
reduced to near zero levels and resultant images show no structural differences when&#xd;
compared to control acquisitions. During gradient application, the CZT modules are&#xd;
suspended, thus yielding a 60% drop in sensitivity due to the additional deadtime of the&#xd;
triggering. The produced MR image showed a reduction in SNR of 92%. Co-registration&#xd;
of the MR and gamma camera images was successfully performed, showing both the&#xd;
physical structural and radioactive activity distribution of the phantom.&#xd;
Additionally, the potential of a diverging fan beam collimator for use in the multimodal SPECT/MR system was investigated using Monte Carlo simulations. Eleven&#xd;
collimator designs were compared, yielding between 13 to 23 modules per ring. The results showed that a practical SPECT/MR design using 18 detectors per ring with a 3.83&#xd;
iii&#xd;
cm length collimator gave similar tomographic resolution to that of a clinical SPECT/CT&#xd;
system but with 7.0 times greater detection sensitivity compared to a conventional rotating dual-head camera. The simulations also showed that the system suffers from aliasing&#xd;
effects when reconstructing features of 7.9 mm or less. To further investigate how the&#xd;
design choices affected the tomographic resolution, parameters for collimator hole size,&#xd;
detector pixel size, and number of projection angles were also explored.</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degreetype" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree" lang="en_US">Master of Science in Biomedical Engineering</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso" lang="en_US">en</dim:field>
   <dim:field mdschema="dc" element="title" lang="en_US">Development of a Cadmium Zinc Telluride Based Gamma Camera for Simultaneous SPECT/MR Imaging</dim:field>
   <dim:field mdschema="dc" element="type" lang="en_US">Thesis</dim:field>
   <dim:field mdschema="others" element="access-status">open.access</dim:field>
</dim:dim></metadata></record></GetRecord></OAI-PMH>