<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-11-21T00:35:07Z</responseDate><request verb="GetRecord" identifier="oai:macsphere.mcmaster.ca:11375/33744" metadataPrefix="dim">https://prod-ms-be.lib.mcmaster.ca/server/oai/request</request><GetRecord><record><header><identifier>oai:macsphere.mcmaster.ca:11375/33744</identifier><datestamp>2026-10-09T07:10:35Z</datestamp><setSpec>com_11375_271</setSpec><setSpec>col_11375_272</setSpec></header><metadata><dim:dim xmlns:dim="http://www.dspace.org/xmlns/dspace/dim" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:doc="http://www.lyncode.com/xoai" xsi:schemaLocation="http://www.dspace.org/xmlns/dspace/dim http://www.dspace.org/schema/dim.xsd">
   <dim:field mdschema="dc" element="contributor" qualifier="advisor">Hildebrandt, Niko</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="author">Switchuk, Marianna</dim:field>
   <dim:field mdschema="dc" element="contributor" qualifier="department">Biomedical Engineering</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="accessioned">2026-10-08T17:53:20Z</dim:field>
   <dim:field mdschema="dc" element="date" qualifier="issued">2026</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://hdl.handle.net/11375/33744</dim:field>
   <dim:field mdschema="dc" element="identifier" qualifier="uri">https://doi.org/10.71548/25514</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="abstract">Accurate and timely measurement of biochemical markers is essential for effective clinical
decision-making, yet many important analytes are still monitored using intermittent laboratory
testing. Potassium, a tightly regulated electrolyte critical for cardiac and neuromuscular
function, can fluctuate rapidly, with both low and high levels posing serious health risks.
Creatinine, an important marker of kidney function, is also typically measured using
centralized laboratory assays that do not support continuous or real-time monitoring. Together,
these limitations highlight an important gap in current diagnostics: the lack of simple, rapid,
and accessible technologies for point-of-care or continuous biomarker monitoring.
This thesis presents the development of a new class of micelle-enhanced optical biosensors
designed to address these challenges. The primary focus is the detection of potassium using a
sensing system based on sodium tetraphenylborate (STB) incorporated within sodium dodecyl
sulfate (SDS) micelles. In this system, potassium induces controlled aggregation within the
micellar environment, producing measurable turbidity changes that correlate with potassium
concentration. Embedding STB within SDS micelles helped reduce the uncontrolled
precipitation observed in conventional systems, improving signal stability, extending the
measurable range, and enabling more reproducible detection in physiologically relevant
matrices such as artificial saliva and interstitial fluid. The platform demonstrated sensitivity
within clinically relevant ranges, with a limit of detection of approximately 0.4 mM and a limit
of quantification of approximately 1.3 mM in artificial saliva.
As a secondary objective, this work also explored the adaptability of the micelle-assisted
sensing strategy for creatinine detection using citrate-capped gold nanoparticles (AuNPs). In
this system, creatinine promotes nanoparticle aggregation, producing a plasmonic red shift that
can be quantified using the absorbance ratio A680/A520. Incorporation of SDS micelles
stabilized the aggregation process, minimized uncontrolled precipitation, and improved signal
reproducibility. The creatinine sensing platform demonstrated detection within physiologically
relevant concentrations, with a limit of detection of approximately 12.3 µM and a limit of
quantification of approximately 41.1 µM in artificial saliva, while showing minimal
interference from common ionic species.
To support future translation beyond proof-of-concept experiments, a portable optical
prototype device was also developed. The prototype integrates a light source, cuvette-based
sample holder, and miniaturized optical detection system capable of rapidly measuring optical
responses in small-volume samples. This demonstrates the feasibility of user-friendly point-of-care biochemical monitoring.
Overall, this work introduces a sensing approach in which micelle-stabilized chemical
interactions transform otherwise unstable aggregation processes into controlled and clinically
useful optical signals. By demonstrating this concept across two clinically important
biomarkers, potassium and creatinine, this research lays the groundwork for future multianalyte sensing platforms with potential applications in non-invasive diagnostics, wearable
monitoring, and personalized healthcare.</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degreetype">Thesis</dim:field>
   <dim:field mdschema="dc" element="description" qualifier="degree">Master of Science (MSc)</dim:field>
   <dim:field mdschema="dc" element="language" qualifier="iso">en</dim:field>
   <dim:field mdschema="dc" element="subject">Potassium</dim:field>
   <dim:field mdschema="dc" element="subject">Creatinine</dim:field>
   <dim:field mdschema="dc" element="subject">Interstitial fluid</dim:field>
   <dim:field mdschema="dc" element="subject">electrochemical sensing</dim:field>
   <dim:field mdschema="dc" element="subject">optical sensing</dim:field>
   <dim:field mdschema="dc" element="subject">point of care testing</dim:field>
   <dim:field mdschema="dc" element="subject">wearable sensors</dim:field>
   <dim:field mdschema="dc" element="subject">micelle-stabilized chemical interactions</dim:field>
   <dim:field mdschema="dc" element="subject">gold nanoparticles</dim:field>
   <dim:field mdschema="dc" element="subject">prototype for ISF sensing</dim:field>
   <dim:field mdschema="dc" element="subject">real-time monitoring</dim:field>
   <dim:field mdschema="dc" element="subject">non-invasive diagnostics</dim:field>
   <dim:field mdschema="dc" element="subject">personalized medicine</dim:field>
   <dim:field mdschema="dc" element="title">Toward Real-Time Potassium and Creatinine Monitoring: Development of an Optical Biosensor for Non-Invasive Detection</dim:field>
   <dim:field mdschema="dc" element="type" lang="en">Thesis</dim:field>
   <dim:field mdschema="others" element="access-status">open.access</dim:field>
</dim:dim></metadata></record></GetRecord></OAI-PMH>